CMS9000 Patient Monitor
Clinic ICU Multipara Patient Monitor
- Model: CMS9000
- Brand: CONTEC
- Certificate: CE FDA
Introduction
Clinic ICU Multipara Patient Monitor can monitor such parameters as ECG, RESP, SpO2, PR, NIBP and dual-channel TEMP, etc. Multipara patient monitor, CMS7000 integrates parameter measurement module, display, and recorder in one device to form compact and portable equipment. At the same time, its built-in replaceable battery provides convenience for patients moving.
Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
- ECG Heart rate (HR)
– ECG waveform
– Arrhythmia and ST-segment analysis - RESP Respiration rate (RR)
– Respiration waveform - SpO2 Pulse oxygen saturation(SpO2)
– Plethysmogram(PLETH) waveform
– Pulse rate(PR)
– Bar graph - NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean Pressure(MEAN)
- TEMP T1, T2, TD
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.
Features
- 12.1” TFT color LCD, 15-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakhstan, Polish, Czech).
- Fanless design is quiet, energy-saving, and clean, which reduces the possibility of cross-infection.
- All-round monitor for adult, pediatric, and neonates.
- With the standard interface, OxyCRG, trend chard, big character interface, and view bed, convenient to observe.
- Waveform, parameter color, and location can be set optionally.
- Operation mode: keys and buttons.
- Display up to 8-channel waveform on one screen.
- Display 7-lead ECG waveform on one screen, and with ECG step function.
- Adopt digital SpO2 technology, anti-motion, and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
- HRV analysis function.
- NIBP measurement mode: Manual/AUTO/Continuous, storage for 4800-group NIBP data.
- Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
- Drug concentration calculation and titration table functions.
- Print trend table data with one key.
- Connect to Central Monitoring System by 3G, WiFi, or wired mode.
- AC/DC, built-in rechargeable lithium battery achieves uninterrupted monitoring.
- Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
Performance
ECG | Lead mode: 3-lead or 5-lead Lead selection: I, II, III, aVR, aVL, aVF, V Waveform: 5-lead, 2-channel 3-lead, 1-channel Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s |
HR: | Measurement and alarm range: 15~350bpm Accuracy: ±1% or ±1bpm, whichever is greater Alarm accuracy: ± 2 bpm Resolution: 1 bpm ST-segment monitoring: Measurement and alarm range: -2.0mV~+2.0mV Accuracy: -0.8mv~+0.8mv ±0.04mv or ±10%, whichever is greater Other range: unspecified Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC Pacemaker: yes |
RESPIRATION | Method: R-F(RA-LL) Impedance Respiration rate: Measurement and alarm range: 0~150rpm Resolution: 1 rpm Measurement accuracy: ±2 rpm Alarm accuracy: ±3 rpm Apnea alarm: 10~40s Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s |
NIBP | Method: Oscillometry Mode: Manual/AUTO/Continuous Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes Measurement period in Continuous mode: 5 minutes Measurement and alarm range: 10 ~ 270 mmHg Resolution: 1 mmHg Cuff pressure accuracy: ±3 mmHg Measurement accuracy: Maximal mean deviation: ±5 mmHg Maximal standard deviation: 8mmHg Over-pressure protection: Adult mode: 297±3 mmHg Pediatric mode: 240±3 mmHg Neonatal mode: 147±3 mmHg |
SpO2 | Measurement and alarm range: 0 ~ 100% Resolution: 1% Measurement accuracy: 70%~100%: ±2%; 0%~69%: unspecified |
PR |
Measurement and alarm range: 30 ~ 250 bpm Measurement accuracy: ±2 bpm or ±2%, whichever is greater |
TEMP | Channel: dual-channel Measurement and alarm range: 0 ~ 50ºC Resolution: 0.1 ºC Accuracy: ±0.1 ºC |
EtCO2 | Method: Sidestream or Mainstream Measurement range for CO2: 0~150mmHg Resolution for CO2: 0 ~ 69 mmHg: 0.1 mmHg 70 ~ 150 mmHg: 0.25 mmHg Accuracy for CO2: 0 ~ 40 mmHg: ±2 mmHg 41 ~ 70 mmHg: ±5% 71 ~ 100 mmHg: ±8% 101 ~ 150 mmHg: ±10% AwRR range: 2~150 rpm AwRR accuracy: ±1 rpm Apnea alarm: yes |
IBP | Channel: dual-channel Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 Measurement and alarm range: -10~300 mmHg Resolution: 1 mmHg Accuracy: ±2% or 1mmHg, whichever is greater Power supply: AC 100-240V, 50/60Hz Safety classification: Class I, type CF defibrillation-proof applied part |
Accessories
- Adult fingertip SpO2 probe (5-pin)
- Adult NIBP cuff
- NIBP extension tube
- ECG lead cable
- ECG electrode
- Temperature probe
- Power cord
- User Manual
Physical characteristic
- Dimension: 31.9 (L) × 16.1 (W) × 26.9 (H) cm
- Weight: 3.6 Kg
Certificate:
CE/FDA
Accessories:
- Temperature Sensor
- SPO2 Sensor
- Adult BP Cuff
- ETCO2 Module
- ECG Cable
- Disposable Sticky Electrodes
Item | No. | Contents | Qty | Remark |
Standard
Configuration |
1 | Main Unit | 1 | |
2 | Power cable | 1 | ||
3 | NIBP Tube | 1 | ||
4 | ECG cable | 1 | adult | |
5 | ECG Electrodes | 10 | ||
6 | SPO2 Sensor | 1 | adult | |
7 | TEMP Sensor | 1 | ||
8 | Cuff | 1 | ||
9 | ETCO2 Module | 1 | ||
10 | User manual | 1 |